Engineering and Computer Science Faculty Publications
In-vitro Performance of Intramedullary Cement Restrictors in Total Hip Arthroplasty
Document Type
Article
Publication Date
6-2003
Journal Title
Journal of Biomechanics
Volume
36
Issue
6
First Page
835
Last Page
843
DOI
10.1016/S0021-9290(03)00017-4
Abstract
Contemporary cementing techniques in total hip arthroplasty include the use of a cement restrictor to occlude the intramedullary canal. As there are many different designs currently available it was the aim of our study to compare the stability of eight different systems. We investigated the displacement and the ability to occlude the femur of these cement restrictors during standardised cementing of artificial and fresh frozen femora. The maximal intramedullary pressures and the displacement of the plugs were continuously recorded and statistically evaluated.
The results revealed significant differences between the tested cement restrictors. The expandable REX™ Cement Stop and the Exeter Plug™ achieved the highest stability and the least cement leakage. The more rigid designs (Palacos Plug®, BUCK™, Universal) in contrast showed inferior performance. Our biomechanical study emphasises the importance of cement restrictor selection, which can have a crucial influence on the fixation of a cemented total hip replacement.
Keywords
Cementing technique; Cement restrictor; Migration; Stability; Total hip arthroplasty
Recommended Citation
Heisel, C.; Norman, Timothy L.; Rupp, R.; Pritsch, M.; Ewerbeck, V.; and Breusch, S. J., "In-vitro Performance of Intramedullary Cement Restrictors in Total Hip Arthroplasty" (2003). Engineering and Computer Science Faculty Publications. 211.
https://digitalcommons.cedarville.edu/engineering_and_computer_science_publications/211